Articolul precedent |
Articolul urmator |
553 0 |
SM ISO690:2012 KOVALENKO, Yu., MIROSHNICHENKO, S., NEVGASYMYI, A., BALASHOV, S., POTRAHOV, N., MIRONOVA, Y.. Low dose radiographic system for pediatrics. In: Nanotechnologies and Biomedical Engineering, Ed. 2, 18-20 aprilie 2013, Chișinău. Technical University of Moldova, 2013, Editia 2, pp. 641-643. ISBN 978-9975-62-343-8.. |
EXPORT metadate: Google Scholar Crossref CERIF DataCite Dublin Core |
Nanotechnologies and Biomedical Engineering Editia 2, 2013 |
||||||
Conferința "International Conference on Nanotechnologies and Biomedical Engineering" 2, Chișinău, Moldova, 18-20 aprilie 2013 | ||||||
|
||||||
Pag. 641-643 | ||||||
|
||||||
Rezumat | ||||||
The paper deals with a specialized digital X-ray system for use in children's trauma centers and maternity hospitals. The basic requirements for such equipment: high contrast and spatial resolution of the images with minimal radiation exposure to the baby. It is proposed the set of equipment which meets these requirements: the microfocus X-ray generator (output power of less than 50 W, the focal spot of less than 50 microns), a special tripod and a digital receiver on a array of the photodiode matrices , which allows to obtain a digital image with a spatial resolution to 9.0 line pairs per millimeter with entrance exposure dose in the plane of the receiver is less than 500 mR |
||||||
Cuvinte-cheie pediatrics, low-dose radiographic system, high contrast and spatial resolution, microdose radiation exposure. |
||||||
|